:MEASure Commands
22
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
467
:MEASure:DEFine
Command Syntax
:MEASure:DEFine <meas_spec>[,<source>]
<meas_spec> ::= {DELay | THResholds}
<source> ::= {CHANnel<n> | FUNCtion<m> | MATH<m> | WMEMory<r>}
<n> ::= 1 to (# analog channels) in NR1 format
<m> ::= 1 to (# math functions) in NR1 format
<r> ::= 1 to (# ref waveforms) in NR1 format
The :MEASure:DEFine command sets up the definition for measurements by
specifying the delta time or threshold values. Changing these values may affect
the results of other measure commands. The table below identifies which
measurement results that can be affected by redefining the DELay specification or
the THResholds values. For example, changing the THResholds definition from the
default 10%, 50%, and 90% values may change the returned measurement result.
:MEASure:DEFine
DELay Command
Syntax
:MEASure:DEFine DELay,<delay spec>[,<source>]
<delay spec> ::= <edge_spec1>,<edge_spec2>
<edge_spec1> ::= [<slope>]<occurrence>
<edge_spec2> ::= [<slope>]<occurrence>
MEASure Command
DELay
THResholds
DUTYcycle
x
DELay
x
x
FALLtime
x
FREQuency
x
NWIDth
x
OVERshoot
x
PERiod
x
PHASe
x
PREShoot
x
PWIDth
x
RISetime
x
VAVerage
x
VRMS
x
Summary of Contents for InfiniiVision 3000T X Series
Page 1: ...Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide...
Page 40: ...40 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 1 What s New...
Page 50: ...50 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 64: ...64 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 254: ...254 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 306: ...306 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 12 DEMO Commands...
Page 364: ...364 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 17 FFT Commands...
Page 442: ...442 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 21 MARKer Commands...
Page 524: ...524 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 22 MEASure Commands...
Page 656: ...656 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 26 POWer Commands...
Page 666: ...666 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 27 RECall Commands...
Page 888: ...888 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 29 SBUS n Commands...
Page 978: ...978 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 30 SEARch Commands...
Page 1240: ...1240 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 38 Error Messages...