822
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
29
:SBUS<n> Commands
:SBUS<n>:SENT:DISPlay
Command Syntax
:SBUS<n>:SENT:DISPlay <base>
<base> ::= {HEX | DECimal | SYMBolic}
The :SBUS<n>:SENT:DISPlay command specifies the number base used by the
decoder. The chosen base is used for the data nibbles in Raw decode format, the
defined Signals in the other formats, and for the data field of the Serial Messages.
This selection is used for both the Lister and the decode line displays.
When SYMBolic is selected, Fast Channel Signals display a calculated physical
value based on the specified multiplier and offset:
• PhysicalValue = (Multiplier * SignalValueAsUnsignedInteger) + Offset
When SYMBolic is selected, the CRC and Slow Channel information is displayed in
hex.
Query Syntax
:SBUS<n>:SENT:DISPlay?
The :SBUS<n>:SENT:DISPlay? query returns the SENT decode number base
setting.
Return Format
<base><NL>
<base> ::= {HEX | DEC | SYMB}
See Also
•
•
•
•
•
•
•
":SBUS<n>:SENT:SIGNal<s>:DISPlay"
•
":SBUS<n>:SENT:SIGNal<s>:LENGth"
•
":SBUS<n>:SENT:SIGNal<s>:MULTiplier"
•
":SBUS<n>:SENT:SIGNal<s>:OFFSet"
•
":SBUS<n>:SENT:SIGNal<s>:ORDer"
•
":SBUS<n>:SENT:SIGNal<s>:STARt"
•
•
•
•
Summary of Contents for InfiniiVision 3000T X Series
Page 1: ...Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide...
Page 40: ...40 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 1 What s New...
Page 50: ...50 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 64: ...64 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 254: ...254 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 306: ...306 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 12 DEMO Commands...
Page 364: ...364 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 17 FFT Commands...
Page 442: ...442 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 21 MARKer Commands...
Page 524: ...524 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 22 MEASure Commands...
Page 656: ...656 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 26 POWer Commands...
Page 666: ...666 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 27 RECall Commands...
Page 888: ...888 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 29 SBUS n Commands...
Page 978: ...978 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 30 SEARch Commands...
Page 1240: ...1240 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 38 Error Messages...