578
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
24
:MTESt Commands
:MTESt:SCALe:XDELta
Command Syntax
:MTESt:SCALe:XDELta <xdelta_value>
<xdelta_value> ::= X delta value in NR3 format
The :MTESt:SCALe:XDELta command defines the position of the X2 marker with
respect to the X1 marker. In the mask test coordinate system, the X1 marker
defines where X=0; thus, the X2 marker defines where X=1.
Because all X vertices of the regions defined for mask testing are normalized with
respect to X1 and
Δ
X, redefining
Δ
X also moves those vertices to stay in the same
locations with respect to X1 and
Δ
X. Thus, in many applications, it is best if you
define XDELta as a pulse width or bit period. Then, a change in data rate without
corresponding changes in the waveform can easily be handled by changing
Δ
X.
The X-coordinate of polygon vertices is normalized using this equation:
X = (X *
Δ
X) + X1
The X delta value is a time value specifying the distance of the X2 marker with
respect to the X1 marker.
For example, if the period of the waveform you wish to test is 1 ms, setting
Δ
X to
1 ms ensures that the waveform's period is between the X1 and X2 markers.
Query Syntax
:MTESt:SCALe:XDELta?
The :MTESt:SCALe:XDELta? query returns the current value of
Δ
X.
Return Format
<xdelta_value><NL>
<xdelta_value> ::= X delta value in NR3 format
See Also
•
"Introduction to :MTESt Commands"
•
•
•
•
Summary of Contents for InfiniiVision 3000T X Series
Page 1: ...Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide...
Page 40: ...40 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 1 What s New...
Page 50: ...50 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 64: ...64 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 254: ...254 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 306: ...306 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 12 DEMO Commands...
Page 364: ...364 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 17 FFT Commands...
Page 442: ...442 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 21 MARKer Commands...
Page 524: ...524 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 22 MEASure Commands...
Page 656: ...656 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 26 POWer Commands...
Page 666: ...666 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 27 RECall Commands...
Page 888: ...888 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 29 SBUS n Commands...
Page 978: ...978 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 30 SEARch Commands...
Page 1240: ...1240 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 38 Error Messages...