:MEASure Commands
22
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
507
:MEASure:TEDGe
Query Syntax
:MEASure:TEDGe? <slope><occurrence>[,<source>]
<slope> ::= direction of the waveform. A rising slope is indicated by a
space or plus sign (+). A falling edge is indicated by a
minus sign (-).
<occurrence> ::= the transition to be reported. If the occurrence number
is one, the first crossing from the left screen edge is
reported.
If the number is two, the second crossing is
reported, etc.
<source> ::= {<digital channels> | CHANnel<n> | FUNCtion<m> | MATH<m>
| WMEMory<r>}
<digital channels> ::= DIGital<d> for the MSO models
<n> ::= 1 to (# of analog channels) in NR1 format
<m> ::= 1 to (# math functions) in NR1 format
<r> ::= 1 to (# ref waveforms) in NR1 format
<d> ::= 0 to (# digital channels - 1) in NR1 format
When the :MEASure:TEDGe query is sent, the displayed signal is searched for the
specified transition. The time interval between the trigger event and this
occurrence is returned as the response to the query. The sign of the slope selects a
rising (+) or falling (-) edge. If no sign is specified for the slope, it is assumed to be
the rising edge.
The magnitude of occurrence defines the occurrence to be reported. For example,
+3 returns the time for the third time the waveform crosses the midpoint threshold
in the positive direction. Once this crossing is found, the oscilloscope reports the
time at that crossing in seconds, with the trigger point (time zero) as the reference.
If the specified crossing cannot be found, the oscilloscope r9.9E+37. This
value is returned if the waveform does not cross the specified vertical value, or if
the waveform does not cross the specified vertical value for the specific number of
times in the direction specified.
You can make delay and phase measurements using the MEASure:TEDGe
command:
Delay = time at the nth rising or falling edge of the channel - time at the same
edge of another channel
Phase = (delay between channels / period of channel) x 360
For an example of making a delay and phase measurement, see
If the optional source parameter is specified, the current source is modified.
Summary of Contents for InfiniiVision 3000T X Series
Page 1: ...Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide...
Page 40: ...40 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 1 What s New...
Page 50: ...50 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 64: ...64 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 254: ...254 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 306: ...306 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 12 DEMO Commands...
Page 364: ...364 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 17 FFT Commands...
Page 442: ...442 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 21 MARKer Commands...
Page 524: ...524 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 22 MEASure Commands...
Page 656: ...656 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 26 POWer Commands...
Page 666: ...666 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 27 RECall Commands...
Page 888: ...888 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 29 SBUS n Commands...
Page 978: ...978 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 30 SEARch Commands...
Page 1240: ...1240 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 38 Error Messages...