:FUNCtion<m> Commands
18
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
389
:FUNCtion<m>:OFFSet
Command Syntax
:FUNCtion<m>:OFFSet <offset>
<m> ::= 1 to (# math functions) in NR1 format
<offset> ::= the value at center screen in NR3 format.
The :FUNCtion<m>:OFFSet command sets the voltage or vertical value
represented at center screen for the selected function. The range of legal values is
gen/-10 times the current scale of the selected function, but will vary by
function. If you set the offset to a value outside of the legal range, the offset value
is automatically set to the nearest legal value.
Query Syntax
:FUNCtion<m>:OFFSet?
The :FUNCtion<m>:OFFSet? query outputs the current offset value for the selected
function.
Return Format
<offset><NL>
<offset> ::= the value at center screen in NR3 format.
See Also
•
"Introduction to :FUNCtion<m> Commands"
•
•
•
NOTE
The :FUNCtion<m>:OFFSet command is equivalent to the :FUNCtion<m>:REFerence
command.
Summary of Contents for InfiniiVision 3000T X Series
Page 1: ...Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide...
Page 40: ...40 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 1 What s New...
Page 50: ...50 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 64: ...64 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 254: ...254 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 306: ...306 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 12 DEMO Commands...
Page 364: ...364 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 17 FFT Commands...
Page 442: ...442 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 21 MARKer Commands...
Page 524: ...524 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 22 MEASure Commands...
Page 656: ...656 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 26 POWer Commands...
Page 666: ...666 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 27 RECall Commands...
Page 888: ...888 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 29 SBUS n Commands...
Page 978: ...978 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 30 SEARch Commands...
Page 1240: ...1240 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 38 Error Messages...