390
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
18
:FUNCtion<m> Commands
:FUNCtion<m>:OPERation
Command Syntax
:FUNCtion<m>:OPERation <operation>
<m> ::= 1 to (# math functions) in NR1 format
<operation> ::= {ADD | SUBTract | MULTiply | DIVide | DIFF | INTegrate
| FFT | SQRT | MAGNify | ABSolute | SQUare | LN | LOG | EXP | TEN
| LOWPass | HIGHpass | AVERage | SMOoth | ENVelope | LINear
| MAXHold | MINHold | TRENd | BTIMing | BSTate}
The :FUNCtion<m>:OPERation command sets the desired waveform math
operator, transform, filter or visualization:
• Operators:
•
ADD — S source2.
•
SUBTract — Source1 - source2.
•
MULTiply — Source1 * source2.
•
DIVide — Source1 / source2.
Operators perform their function on two analog channel sources.
• Transforms:
•
DIFF — Differentiate
•
INTegrate — The INTegrate:IOFFset command lets you specify a DC offset
correction factor.
•
FFT — The SPAN, CENTer, VTYPe, and WINDow commands are used for FFT
functions. When FFT is selected, the horizontal cursors change from time to
frequency (Hz), and the vertical cursors change from volts to decibel (dB).
•
LINear — Ax + B — The LINear commands set the gain (A) and offset (B)
values for this function.
•
SQUare
•
SQRT — Square root
•
ABSolute — Absolute Value
•
LOG — Common Logarithm
•
LN — Natural Logarithm
•
EXP — Exponential (e
x
)
•
TEN — Base 10 exponential (10
x
)
Transforms operate on a single analog channel source or on lower math
functions.
• Filters:
•
LOWPass — Low pass filter — The FREQuency:LOWPass command sets the
-3 dB cutoff frequency.
Summary of Contents for InfiniiVision 3000T X Series
Page 1: ...Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide...
Page 40: ...40 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 1 What s New...
Page 50: ...50 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 64: ...64 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 254: ...254 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 306: ...306 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 12 DEMO Commands...
Page 364: ...364 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 17 FFT Commands...
Page 442: ...442 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 21 MARKer Commands...
Page 524: ...524 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 22 MEASure Commands...
Page 656: ...656 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 26 POWer Commands...
Page 666: ...666 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 27 RECall Commands...
Page 888: ...888 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 29 SBUS n Commands...
Page 978: ...978 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 30 SEARch Commands...
Page 1240: ...1240 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 38 Error Messages...