1152
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
35
:WGEN<w> Commands
:WGEN<w>:MODulation:NOISe
Command Syntax
:WGEN<w>:MODulation:NOISe <percent>
<w> ::= 1 or 2 in NR1 format
<percent> ::= 0 to 100
The :WGEN<w>:MODulation:NOISe command adds noise to the currently selected
signal. The sum of the amplitude between the original signal and injected noise is
limited to the regular amplitude limit (for example, 5 Vpp in 1 MOhm), so the range
for <percent> varies according to current amplitude.
Note that adding noise affects edge triggering on the waveform generator source
as well as the waveform generator sync pulse output signal (which can be sent to
TRIG OUT). This is because the trigger comparator is located after the noise
source.
Query Syntax
:WGEN<w>:MODulation:NOISe?
The :WGEN<w>:MODulation:NOISe query returns the percent of added noise.
Return Format
<percent><NL>
<percent> ::= 0 to 100
See Also
•
Summary of Contents for InfiniiVision 3000T X Series
Page 1: ...Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide...
Page 40: ...40 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 1 What s New...
Page 50: ...50 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 64: ...64 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 254: ...254 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 306: ...306 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 12 DEMO Commands...
Page 364: ...364 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 17 FFT Commands...
Page 442: ...442 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 21 MARKer Commands...
Page 524: ...524 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 22 MEASure Commands...
Page 656: ...656 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 26 POWer Commands...
Page 666: ...666 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 27 RECall Commands...
Page 888: ...888 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 29 SBUS n Commands...
Page 978: ...978 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 30 SEARch Commands...
Page 1240: ...1240 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 38 Error Messages...