Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
3
Getting Started
When you send the :DIGitize command to the oscilloscope, the specified channel
signal is digitized with the current :ACQuire parameters. To obtain waveform data,
you must specify the :WAVeform parameters for the SOURce channel, the FORMat
type, and the number of POINts prior to sending the :WAVeform:DATA? query.
The number of data points comprising a waveform varies according to the number
requested in the :ACQuire subsystem. The :ACQuire subsystem determines the
number of data points, type of acquisition, and number of averages used by the
:DIGitize command. This allows you to specify exactly what the digitized
information contains.
The following program example shows a typical setup:
myScope.WriteString ":ACQuire:TYPE AVERage"
myScope.WriteString ":ACQuire:COMPlete 100"
myScope.WriteString ":ACQuire:COUNt 8"
myScope.WriteString ":DIGitize CHANnel1"
myScope.WriteString ":WAVeform:SOURce CHANnel1"
myScope.WriteString ":WAVeform:FORMat BYTE"
myScope.WriteString ":WAVeform:POINts 500"
myScope.WriteString ":WAVeform:DATA?"
This setup places the instrument into the averaged mode with eight averages. This
means that when the :DIGitize command is received, the command will execute
until the signal has been averaged at least eight times.
After receiving the :WAVeform:DATA? query, the instrument will start passing the
waveform information.
Digitized waveforms are passed from the instrument to the controller by sending a
numerical representation of each digitized point. The format of the numerical
representation is controlled with the :WAVeform:FORMat command and may be
selected as BYTE, WORD, or ASCii.
NOTE
Ensure New Data is Collected
When you change the oscilloscope configuration, the waveform buffers are cleared. Before
doing a measurement, send the :DIGitize command to the oscilloscope to ensure new data has
been collected.
NOTE
Set :TIMebase:MODE to MAIN when using :DIGitize
:TIMebase:MODE must be set to MAIN to perform a :DIGitize command or to perform any
:WAVeform subsystem query. A "Settings conflict" error message will be returned if these
commands are executed when MODE is set to ROLL, XY, or WINDow (zoomed). Sending the
*RST (reset) command will also set the time base mode to normal.
Summary of Contents for InfiniiVision 3000T X Series
Page 1: ...Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide...
Page 40: ...40 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 1 What s New...
Page 50: ...50 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 64: ...64 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 254: ...254 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 306: ...306 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 12 DEMO Commands...
Page 364: ...364 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 17 FFT Commands...
Page 442: ...442 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 21 MARKer Commands...
Page 524: ...524 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 22 MEASure Commands...
Page 656: ...656 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 26 POWer Commands...
Page 666: ...666 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 27 RECall Commands...
Page 888: ...888 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 29 SBUS n Commands...
Page 978: ...978 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 30 SEARch Commands...
Page 1240: ...1240 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 38 Error Messages...