796
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
29
:SBUS<n> Commands
:SBUS<n>:LIN Commands
NOTE
These commands are valid when the automotive CAN and LIN serial decode option (Option
AMS) has been licensed.
Table 113
:SBUS<n>:LIN Commands Summary
Command
Query
Options and Query Returns
:SBUS<n>:LIN:PARity
{{0 | OFF} | {1 |
ON}} (see
:SBUS<n>:LIN:PARity?
{0 | 1}
:SBUS<n>:LIN:SAMPlepo
int <value> (see
:SBUS<n>:LIN:SAMPlepo
<value> ::= {60 | 62.5 | 68 | 70
| 75 | 80 | 87.5} in NR3 format
:SBUS<n>:LIN:SIGNal:B
AUDrate <baudrate>
(see
:SBUS<n>:LIN:SIGNal:B
AUDrate? (see
<baudrate> ::= integer from 2400
to 625000 in 100 b/s increments
:SBUS<n>:LIN:SOURce
<source> (see
:SBUS<n>:LIN:SOURce?
<source> ::= {CHANnel<n> |
EXTernal} for DSO models
<source> ::= {CHANnel<n> |
DIGital<d>} for MSO models
<n> ::= 1 to (# analog channels)
in NR1 format
<d> ::= 0 to (# digital channels
- 1) in NR1 format
:SBUS<n>:LIN:STANdard
<std> (see
:SBUS<n>:LIN:STANdard
<std> ::= {LIN13 | LIN20}
:SBUS<n>:LIN:SYNCbrea
k <value> (see
:SBUS<n>:LIN:SYNCbrea
<value> ::= integer = {11 | 12 |
13}
:SBUS<n>:LIN:TRIGger
<condition> (see
:SBUS<n>:LIN:TRIGger?
<condition> ::= {SYNCbreak | ID |
DATA}
:SBUS<n>:LIN:TRIGger:
ID <value> (see
:SBUS<n>:LIN:TRIGger:
<value> ::= 7-bit integer in
decimal, <nondecimal>, or
<string> from 0-63 or 0x00-0x3f
<nondecimal> ::= #Hnn where n ::=
{0,..,9 | A,..,F} for hexadecimal
<nondecimal> ::= #Bnn...n where n
::= {0 | 1} for binary
<string> ::= "0xnn" where n ::=
{0,..,9 | A,..,F} for hexadecimal
Summary of Contents for InfiniiVision 3000T X Series
Page 1: ...Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide...
Page 40: ...40 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 1 What s New...
Page 50: ...50 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 64: ...64 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 254: ...254 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 306: ...306 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 12 DEMO Commands...
Page 364: ...364 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 17 FFT Commands...
Page 442: ...442 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 21 MARKer Commands...
Page 524: ...524 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 22 MEASure Commands...
Page 656: ...656 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 26 POWer Commands...
Page 666: ...666 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 27 RECall Commands...
Page 888: ...888 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 29 SBUS n Commands...
Page 978: ...978 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 30 SEARch Commands...
Page 1240: ...1240 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 38 Error Messages...