:MARKer Commands
21
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
433
:MARKer:X2Y2source
Command Syntax
:MARKer:X2Y2source <source>
<source> ::= {CHANnel<n> | FUNCtion<m> | MATH<m> | WMEMory<r>}
<n> ::= 1 to (# analog channels) in NR1 format
<m> ::= 1 to (# math functions) in NR1 format
<r> ::= 1 to (# ref waveforms) in NR1 format
The :MARKer:X2Y2source command sets the source for the cursors. The channel
you specify must be enabled for cursors to be displayed. If the channel or function
is not on, an error message is issued.
If the marker mode is not currently WAVeform (see
page 429):
• Sending a :MARKer:X2Y2source command will put the cursors in the MANual
mode.
• Setting the source for one pair of markers (for example, X2Y2) sets the source
for the other (for example, X1Y1).
If the marker mode is currently WAVeform, the X2Y2 source can be set separate
from the X1Y1 source.
If :MARKer:MODE is set to OFF or MANual, setting :MEASure:SOURce to
CHANnel<n>, FUNCtion, MATH, or WMEMory<r> will also set
:MARKer:X1Y1source and :MARKer:X2Y2source to this value.
Query Syntax
:MARKer:X2Y2source?
The :MARKer:X2Y2source? query returns the current source for the cursors. If all
channels are off or if :MARKer:MODE is set to OFF, the query returns NONE.
Return Format
<source><NL>
<source> ::= {CHAN<n> | FUNC | WMEM<r> | NONE}
See Also
•
"Introduction to :MARKer Commands"
•
•
•
NOTE
MATH is an alias for FUNCtion. The query will return FUNC if the source is FUNCtion or MATH.
Summary of Contents for InfiniiVision 3000T X Series
Page 1: ...Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide...
Page 40: ...40 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 1 What s New...
Page 50: ...50 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 64: ...64 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 254: ...254 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 306: ...306 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 12 DEMO Commands...
Page 364: ...364 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 17 FFT Commands...
Page 442: ...442 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 21 MARKer Commands...
Page 524: ...524 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 22 MEASure Commands...
Page 656: ...656 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 26 POWer Commands...
Page 666: ...666 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 27 RECall Commands...
Page 888: ...888 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 29 SBUS n Commands...
Page 978: ...978 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 30 SEARch Commands...
Page 1240: ...1240 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 38 Error Messages...