INFICON
Description | 4
Pernicka-700H-CHLD-Operating-instructions-jin76aen1-04-(1911)
27 / 100
Pressure supply purge gas (Argon 5.0)
35 mbar - 70 mbar (over-pressure)
Vacuum system
Cryogenic pump
Turbo molecular pump
Backing pump
Detectable masses
2 - 100
Connection of customer test leak
DN 16 CF
4.3.6.4 Ambient conditions
Device
Permissible ambient temperature (during
operation)
15 °C ... 28 °C
Permissible storage temperature
10 °C ... 45 °C
Max. relative humidity up to 31 °C
80 %
Max. relative humidity from 31 °C to 40
°C
linearly decreasing from 80 % to 50 %
Max. relative humidity above 40 °C
50 %
Pollution degree
II
Max. altitude above sea level
1000 m
Summary of Contents for Pernicka 700H CHLD