6 | Operation
INFICON
68 / 100
Pernicka-700H-CHLD-Operating-instructions-jin76aen1-04-(1911)
The "MS-Panel" window displays the measured data during a measurement process.
The corresponding 10 measured values recorded during an accumulation phase for
the different masses are documented.
6.5.4.3 Display current settings
► To open the "System" window, press the tab "System".
Fig. 40: Window "System"
The current settings for the completed measurement process are displayed in the
"System" window. (Pin 17)
Output parameter/
Pin assignment of
the digital I/O *)
Status
Ready (Pin 17)
Ready = HIGH: You can start a measurement / calibration.
Pass (Pin 18)
Pass = HIGH: The tested test object is leak-proof.
Fine Leak (Pin 19)
Gross Leak (Pin 21)
Fine Leak = HIGH and Gross Leak = LOW: Helium fine leak.
Fine Leak = LOW and Gross Leak = HIGH: Helium gross leak.
Fine Leak = HIGH and Gross Leak = HIGH: Gas-2-fine leak.
Error (Pin 20)
Error = HIGH: Error message displayed on the screen.
*) The bushing of the CHLD detector is equipped with a plug by default. If you need to
use the output signals, use a Y-cable.
Further explanations: The outputs are open collectors, max. 60 mA per output, Pin 10
is +24 V, Pin 9 is 0 V.
6.5.4.4 Display general information
► To open the "Locals" window, press the tab "Locals".