6 | Operation
INFICON
48 / 100
Pernicka-700H-CHLD-Operating-instructions-jin76aen1-04-(1911)
MSC: The MSC (Mass Scale Correction) function is not used.
hide: To hide the tabs from "Mixed Data".
► To display all tabs from "Mixed Data" in the header, which can be used to open the
same window with additional detail information, deactivate this option. This function
is protected by the password "Germany".
TipStrip: For displaying an information text, when the mouse pointer points to an
operating element, for example, a button for a certain time, for which a corresponding
information text has been programmed.
Helium calibration leak
To adjust the parameters of the helium test leak.
► After a changed helium calibration leak, reset the default parameters according to
the supplied certificate.
AutoStartUp
Auto Filament: Activate the cathode in the CHLD-Sensor automatically after
evacuation.
► To enable the cathode in the CHLD-Sensor automatically after a completed
evacuation, activate this option.
Valve Control
Valve Delay: To set the valve delay time. This delay time ensures that two valves do
not change their state at the same time.
► Set the value "3000", i.e. a time delay of 3 seconds between the respective start of
two valve movements.
User Form
Print User Form: To print the measurement data after the measurement process.
► To print the measurement data after each measurement process, activate this
option.
Clear Serial Number: To reset the serial number entered in the "Input" window after a
measurement.
► To automatically delete the serial number of the measured object after each
measurement process, activate this option.
LR unit: To select the unit of measure for the leak rate.
► Choose between the unit of measurement [mbar l/s] or [atm cc/s].
Load/Unload Time [s]
Duration to purge the test chamber.
Summary of Contents for Pernicka 700H CHLD