5.3 Connecting the compressor
NOTICE
Damage to the cryogenic pump compressor due to incorrect settings
Incorrect setting of mains voltage and mains frequency can damage the cryogenic
pump compressor.
To correctly set the cryogenic pump compressor on the device with the order number
550-700, set the cryogenic pump compressor to: 208 V, 60 Hz.
To correctly set the cryogenic pump compressor on the device with the order number
550-701, set the cryogenic pump compressor to: 230 V, 50 Hz.
NOTICE
Damage to the cryogenic pump compressor due to unfavorable installation
location
Too small and not sufficiently ventilated location can damage the cryogenic pump
compressor.
► Choose as a location for the compressor a sufficiently ventilated room and a
minimum distance of the compressor to the walls of 1 m.
Cryogenic pump compressor
Read and follow the included documentation "Cryo-Torr Pump Installation, Operation
and Maintenance Instructions".