6 | Operation
INFICON
72 / 100
Pernicka-700H-CHLD-Operating-instructions-jin76aen1-04-(1911)
6.8 Perform regeneration
On a daily basis, before switching off the device and mainly after a series of
successive measurements, the gases frozen in the cryogenic pump must be
evaporated and removed from the system.
► Perform daily regeneration of the cryogenic pump.
After the regeneration is started, the heating phase is started after a pre-warning
period. The temperature in the cryogenic pump increases and the frozen gases are
evaporated and pumped out, depending on their dew point.
1
Mini > Regenerate or Measure > Regenerate.
Button "
Button "Set Time to Now
Set Time and Date "
"
Fig. 46: Input window "Regeneration cryogenic pump"
2
To define the regeneration period, choose between the 2 possibilities.
ð
To start regeneration on a daily basis with set start time, select "weekday". To
set the start time with the "Start-Time" parameter the end time with the "End-
Time" parameter press the "Set Time and Date" button. To start the
regeneration immediately, press the "Set Time to Now" button.
ð
To start regeneration before each weekend with set start time, select
"weekend". To set the start time with the "Start-Time" parameter the end time
with the "End-Time" parameter press the "Set Time and Date" button. To start
the regeneration immediately, press the "Set Time to Now" button.
3
To start the regeneration at the preset start time, press the "OK" button.
ð
The measuring program automatically changes to the "Regeneration
cryogenic pump" window.