Fig. 47: "Regeneration cryogenic pump" window
The "Regeneration cryogenic pump" window gives you a detailed overview of the
regeneration history.
To stop regeneration before the set end time, press the "Abort" button.
6.9 Switching off the device.
NOTICE
Damage to the cryogenic pump
If there are frozen gases in the cryogenic pump and the device is switched off, the
gases are evaporated and the rupture disk on the cryogenic pump can rupture as a
result of the rising pressure.
► Perform a regeneration of the cryogenic pump, at least up to a temperature of 280
K, before you switch off the device.
You can turn the device off with the power switch on the back of the device.
The parameters set in the device remain saved.
1
Stop the Pernicka 700H software.
2
Shut down the computer.
3
To turn the device off, set the power switch to OFF. When the device is switched
off, the compressor of the cryogenic pump is also switched off.