► To define the duration in which the test chamber is purged with purge gas, set the
time.
Short regen after Gross Leak
Short-term heating of the cryogenic pump after a gross leak test to remove helium
from the cryogenic pump. This function is activated by default.
After a gross leak test, helium atoms are located between the purge gas atoms on the
surface of the cryogenic pump. In order to avoid increased leakage rates during
subsequent measurements by means of this helium fraction, the cryogenic pump is
heated up briefly, but only when during the gross leak test a gross leak is detected, so
that the helium components can be evaporated and removed from the system.
• To turn on this feature, click the icon. The triangle in the symbol is displayed in
green.
• To turn off this feature, click the icon. The triangle in the symbol is displayed in
black.
Use User Input Form
"Input" window for detailed information on the test object.
• To enter only the serial number for each test object, deselect the "Use User Input
Form" option.
• To enter detailed information for each test object, select the "Use User Input Form"
option. The "Input" window opens.
6.3 Select method
The settings and configurations required for measuring the leak rate are summarized
in the method editor in a so-called method. Assign a corresponding method before
beginning each individual or serial measurement.
You can use preset methods or create new methods yourself.
6.3.1 Select an existing method
Using a method, you define the times and limits to be set over the measurement
sequence of a leakage measurement.
1
Mini > Method > Load & Run or Measure > Method > Load & Run.