6 | Operation
INFICON
42 / 100
Pernicka-700H-CHLD-Operating-instructions-jin76aen1-04-(1911)
If you cannot start the device without entering a password …
In case of an accidental closing of the user login, it may happen that you are
requested to enter a password.
► Do not change the standard user and enter the password “Yes“.
6.2 Basic settings
6.2.1 Setting the measurement window
For operating and configuring the device in the “Measurement” operating mode, you
can choose between two windows.
6.2.1.1 Set the "Mini" window
► To open the "Mini” window, press the tab "Mini".
Display of the leak rate
Message window
Display
measuring progress
Amplifier
ON/OFF
Cathode
in
Method in use
CHLD-Sensor
ON/OFF
SEM -
Fig. 20: Window "Mini"
The "Mini" window offers all the settings required for the measurement process and
can be used for serial measurements of test objects.
This window is mainly suitable for the operator of the device.
Button
Description
Load Method
► To start the method editor , press this button. See