5.8 Mounting the test chamber
► To mount or to change the test chamber, place the device in the "Standby"
operating mode and then turn off the device. The device is in the "Standby"
operating mode before a measurement process is started and after a completed
measurement process.
► To prevent grease deposits on the device, use neoprene or nitrile disposable gloves
free of powder during operation.
Accessories required
Test chamber small, medium or large
Required tools
• Allen key, modified, 9/64 inch (included)
• General purpose pliers of small or slip joint pliers
Test chamber
Screw connection
purge gas
supply
Flange with
screws
Purge gas
supply
Fig. 15: Test chamber: built-in state
1
If the inlet flange is covered by a dummy flange (normal newly delivered
condition), you have to remove the 6 screws of the flange connection and take
out the dummy flange.
2
If a test chamber is mounted on the inlet flange, loosen the screws from the
purge gas supply and remove it from the test chamber.
Loosen and remove the 6 screws of the flange connection and carefully remove
the test chamber.