4 | Description
INFICON
26 / 100
Pernicka-700H-CHLD-Operating-instructions-jin76aen1-04-(1911)
4.3.6 Technical data
4.3.6.1 Mechanical data
Device
Dimensions (W x H x D)
660 mm x 1390 mm x 870 mm
Weight
245 kg
4.3.6.2 Electrical data
Device
Main fuse
Internal over-current protection,
resettable
Nominal power
2300 VA
Nominal voltage
110 V, 50/60 Hz (Order number:
550-700)
230 V, 50/60 Hz (Order number:
550-701)
Nominal current
15 A (Order number: 550-700)
10 A (Order number: 550-701)
Protection class
IP 20
Excess voltage category
II
Electronic interfaces
USB, Ethernet interface
Compressor for cryogenic pump
Nominal voltage
208 V ... 240 V, 50/60 Hz
Nominal current
10 A
4.3.6.3 Physical data
Device
Minimum detectable leak rate for helium
> 4 x 10
-14
mbar l/s
Leak rate internal test leak
approx. 5 x 10
-10
mbar l/s
Pneumatic pressure valves
7.5 bar - 9 bar
Pressure of the helium supply for the
cryogenic pump compressor
16.9 bar - 17.25 bar (over-pressure)