Installation | 5
Pernicka-700H-CHLD-Operating-instructions-jin76aen1-04-(1911)
37 / 100
2
Plug the supplied power cord into the MAINS CRYO COMPRESSOR plug and
connect the power cord to the power supply.
3
Plug the supplied power cord into the POWER LINE INPUT plug and connect
the power cord to the power supply.
5.7 Connecting the device to an exhaust system
(optional)
1
To turn the device off, set the power switch to OFF. Disconnect the device from
the power supply and secure the device against re-activation.
2
Switch off the backing pump.
3
Hold the exhaust filter with one hand and loosen the wing nut of the clamping
ring.
4
Remove the exhaust filter from the backing pump.
5
Use a suitable connection piece, e.g. a 90-degree hose adapter with KF 16 on
the pump side, onto the backing pump.
6
Tighten the wing nut of the clamping ring.
7
Connect the connecting piece to your exhaust system.
Exhaust filter
Clamping ring
Backing pump
Backing pump
Wing nut
Fig. 14: Connecting the device to an exhaust system