5 | Installation
INFICON
36 / 100
Pernicka-700H-CHLD-Operating-instructions-jin76aen1-04-(1911)
5.6 Connect the device to the mains power
WARNING
Danger from electric shock
Improperly earthed or protected products may be dangerous to life in case of a fault.
The use of the device is not permitted without a connected protective conductor.
► Only use the included 3-wire power cord provided.
CAUTION
Danger due to incorrect mains voltage and mains frequency
Incorrect mains voltage and incorrect mains frequency can lead to the destruction of
the device and personal injury.
► Check that the mains voltage and mains frequency permissible for the Pernicka
700H CHLD as well as the included cryogenic pump compressor comply with the
mains voltage and mains frequency available at the place of use.
Protective conductor
connection point
Connection for power supply
cryogenic pump compressor
Connection for the device power supply
Fig. 13: Connect the device to the mains power
1
Install the protective conductor with a sufficient length and cross-section to the
protective conductor connection point of the device.