8 | Cleaning and maintenance
INFICON
82 / 100
Pernicka-700H-CHLD-Operating-instructions-jin76aen1-04-(1911)
3
Place the QME 220 electronics device on a suitable base that allows you to
work comfortably on the front panel.
Screw
Screw (hidden)
Electronic device
Analyzer
QMA 220
QME 220
Fig. 52: CHLD-Sensor: Unscrew the QME 220 electronics device
Check or change the filter mat
1
To remove the IO 220 module, loosen the 3 screws.
Screw
Screw
Screw
Torx T10
Torx
Torx T10
T10
Fig. 53: CHLD-Sensor: Remove the IO 220 module