Pernicka-700H-CHLD-Operating-instructions-jin76aen1-04-(1911)
25 / 100
4.3.4.1 Adjust the filtering of the measuring signals
INFICON GmbH recommends that you do not change the factory settings.
Fig. 7: Window "IO 220"
► To open the "IO 220” window, press the tab "IO 220".
The active digital inputs and outputs are displayed in the "IO 220" window. The same
information is also provided for the analog signals. The upper window shows the raw
data of the analog signals. The lower window shows the analog signals with the
selected filter.
1
To influence the filtering, select the corresponding values in the field "AI
Coefficients".
2
Press the "reset/update" button.
ð
The result of the changed filter is shown directly in the lower window.
4.3.5 Markings on the device
The markings on the device have the following meanings:
Operate the device only with the protective conductor (PE) connected.
Only operate the device with the protective conductor cable connected to
the electronics box.