6 Operation
6.1 Switching the device on
NOTICE
Damage to the cryogenic pump compressor due to incorrect settings
Incorrect setting of mains voltage and mains frequency can damage the cryogenic
pump compressor.
To correctly set the cryogenic pump compressor on the device with the order number
550-700, set the cryogenic pump compressor to: 208 V, 60 Hz.
To correctly set the cryogenic pump compressor on the device with the order number
550-701, set the cryogenic pump compressor to: 230 V, 50 Hz.
Cryogenic pump compressor
Read and follow the included documentation "Cryo-Torr Pump Installation, Operation
and Maintenance Instructions".
1
Ensure that the correct mains voltage and the correct mains frequency are set
on the compressor.
2
To turn the unit on, set the power switch to ON.
ð
The device has a power-on delay and is switched on after approx. 20 s.
3
To turn on the compressor of the cryogenic pump, set the power switch on the
compressor to ON.
ð
When delivered the device shows the screen "Measure" after starting. See
"Setting the "Measure" window [
44]".
Fig. 19: Window "Measure"