Installation | 5
Pernicka-700H-CHLD-Operating-instructions-jin76aen1-04-(1911)
29 / 100
Present
Not
present
Purge gas supply: Argon 5.0 or better
Alternative purge gases are permissible depending on
the application. Please contact INFICON GmbH.
⃞
⃞
Pressure reducing valve for purge gas supply: < 50 mbar
rel., 10 l/min.
Recommendation: Use the INFICON pressure reducer,
catalog number: 551-700S, available for 4 different bottle
threads
⃞
⃞
Purge gas supply from the pressure reducer to the
device:
6.35 mm copper or stainless steel pipe
⃞
⃞
Helium background in the operating environment of the
device: < 6 ppm
⃞
⃞
► Install the device, the cryogenic pump compressor and all parts included in the
scope of delivery. See original operating instructions, chapter "Installation [
5.2 Setup
DANGER
Health risk due to exhaust fumes and vapors
Exhaust fumes and vapors from pumps can be a health risk.
► When operating in badly vented rooms connect an exhaust hose onto the exhaust
connection when using gases which cause a health risk (see "Connecting the
device to an exhaust system (optional) [
37]").