ET User’s Guide
3-45
Making Measurements
Measuring Conversion Loss
Measuring Conversion Loss
Conversion loss is the ratio of IF output power to RF input power
expressed in dB. This section uses an example measurement to describe
how to measure the conversion loss of a broadband mixer.
When characterizing a device's conversion loss, the analyzer uses
broadband detection to compare the transmitted signal (B*) to the
reference signal (R*). This is because the input and output signals of a
frequency-translating device may be different. Since broadband
detection measures signals at all frequencies, you may want to use a
filter to remove unwanted signals such as LO feedthrough when
performing this measurement.
For example, an RF signal at 900 MHz mixed with an LO signal at
200 MHz, results in mixing products at 700 MHz and 1100 MHz, as well
as the original 900 MHz and 200 MHz RF and LO signals.
Figure 3-21
Filtering Out the Unwanted Mixing Products
Summary of Contents for HP 8712ET
Page 8: ...viii ET User s Guide ...
Page 17: ...ET User s Guide 1 1 1 Installing the Analyzer ...
Page 34: ...1 18 ET User s Guide Installing the Analyzer Preventive Maintenance ...
Page 35: ...ET User s Guide 2 1 2 Getting Started ...
Page 52: ...2 18 ET User s Guide Getting Started Performing the Operator s Check ...
Page 53: ...ET User s Guide 3 1 3 Making Measurements ...
Page 118: ...3 66 ET User s Guide Making Measurements Measuring Impedance Magnitude ...
Page 119: ...ET User s Guide 4 1 4 Using Instrument Functions ...
Page 143: ...ET User s Guide 4 25 Using Instrument Functions Using Markers Figure 4 15 Delta Marker Mode ...
Page 149: ...ET User s Guide 4 31 Using Instrument Functions Using Limit Testing Figure 4 16 Limit Lines ...
Page 207: ...ET User s Guide 5 1 5 Optimizing Measurements ...
Page 226: ...5 20 ET User s Guide Optimizing Measurements Measuring Devices with Long Electrical Delay ...
Page 227: ...ET User s Guide 6 1 6 Calibrating for Increased Measurement Accuracy ...
Page 266: ...6 40 ET User s Guide Calibrating for Increased Measurement Accuracy Checking the Calibration ...
Page 267: ...ET User s Guide 7 1 7 Front Rear Panel ...
Page 292: ...7 26 ET User s Guide Front Rear Panel Line Module ...
Page 293: ...ET User s Guide 8 1 8 Hardkey Softkey Reference ...
Page 395: ...ET User s Guide 9 1 9 Specifications ...
Page 397: ...ET User s Guide 9 3 Specifications System Performance ...
Page 443: ...ET User s Guide 10 1 10 Safety and Regulatory Information ...
Page 449: ...ET User s Guide 10 7 Safety and Regulatory Information Regulatory Information ...
Page 450: ...10 8 ET User s Guide Safety and Regulatory Information Regulatory Information ...
Page 451: ...ET User s Guide 11 1 11 Factory Preset State and Memory Allocation ...