
3-22
ET User’s Guide
Making Measurements
Measuring Transmission Response
Calibrating for Noninsertable Devices
A noninsertable device is one that cannot be inserted in place of a
zero-length through-standard. It has the same connectors on each port
(type and sex) or has a different type of connector on each port (SMA on
one port, and type-N on the other, for example). The following two
calibration methods are available for noninsertable devices:
• swap equal adapters
• modify the calibration kit definition of the through standard
In the following examples, the noninsertable device is type-N, with
female input and output connectors.
Method A: Swap Equal Adapters.
With this method, you use two precision matched adapters which are
“equal.” To be equal, the adapters must have the same match,
characteristic impedance (Z
0
), insertion loss, and electrical delay. The
adapters in most HP calibration kits have matched electrical lengths,
even if the physical lengths appear different.
NOTE
For analyzers with 50 ohm input impedance only: in the 50 ohm type-N
calibration kit (HP 85032B), there are four equal adapters: two
APC-7 to type-N(f), and two APC-7 to type-N(m). To create adapter A in
the following example, connect an APC-7-to-type-N(f) adapter with an
APC-7 to type-N(m) adapter. To create adapter B, connect two
APC-7 to type-N(m) adapters.
NOTE
By convention, cal kit labels in the analyzer indicate the sex of the port
with which they are used. For example, the default cal kit for the
analyzer is type-N female because the front panel RF ports are female
(the calibration standards, in turn, have male connectors).
1. Connect a test cable with type-N(m) connectors to the
TRANSMISSION port.
2. To select a calibration kit for the REFLECTION port, press
, and use the front panel knob or the
keys to highlight the REFLECTION port in the display’s table.
Press
.
3. Connect adapter A (type-N(m) to type-N(f)) to the REFLECTION
port.
CAL
More Cal Cal Kit
Type-N(m)
Summary of Contents for HP 8712ET
Page 8: ...viii ET User s Guide ...
Page 17: ...ET User s Guide 1 1 1 Installing the Analyzer ...
Page 34: ...1 18 ET User s Guide Installing the Analyzer Preventive Maintenance ...
Page 35: ...ET User s Guide 2 1 2 Getting Started ...
Page 52: ...2 18 ET User s Guide Getting Started Performing the Operator s Check ...
Page 53: ...ET User s Guide 3 1 3 Making Measurements ...
Page 118: ...3 66 ET User s Guide Making Measurements Measuring Impedance Magnitude ...
Page 119: ...ET User s Guide 4 1 4 Using Instrument Functions ...
Page 143: ...ET User s Guide 4 25 Using Instrument Functions Using Markers Figure 4 15 Delta Marker Mode ...
Page 149: ...ET User s Guide 4 31 Using Instrument Functions Using Limit Testing Figure 4 16 Limit Lines ...
Page 207: ...ET User s Guide 5 1 5 Optimizing Measurements ...
Page 226: ...5 20 ET User s Guide Optimizing Measurements Measuring Devices with Long Electrical Delay ...
Page 227: ...ET User s Guide 6 1 6 Calibrating for Increased Measurement Accuracy ...
Page 266: ...6 40 ET User s Guide Calibrating for Increased Measurement Accuracy Checking the Calibration ...
Page 267: ...ET User s Guide 7 1 7 Front Rear Panel ...
Page 292: ...7 26 ET User s Guide Front Rear Panel Line Module ...
Page 293: ...ET User s Guide 8 1 8 Hardkey Softkey Reference ...
Page 395: ...ET User s Guide 9 1 9 Specifications ...
Page 397: ...ET User s Guide 9 3 Specifications System Performance ...
Page 443: ...ET User s Guide 10 1 10 Safety and Regulatory Information ...
Page 449: ...ET User s Guide 10 7 Safety and Regulatory Information Regulatory Information ...
Page 450: ...10 8 ET User s Guide Safety and Regulatory Information Regulatory Information ...
Page 451: ...ET User s Guide 11 1 11 Factory Preset State and Memory Allocation ...