3-34
ET User’s Guide
Making Measurements
Measuring Reflection Response
NOTE
By convention, cal kit labels in the analyzer indicate the sex of the port
with which they are used. For example, the default cal kit for the
analyzer is type-N female because the front panel RF ports are female
(the calibration standards, in turn, have male connectors).
NOTE
If you are going to be using calibration standards other than the default
(female type-N), you must select the connector type by pressing
and then selecting the appropriate type.
provides detail about when this calibration is necessary. If you
wish to perform a one-port reflection calibration on your instrument for a
reflection response measurement, execute the following steps:
1. Press
.
2. The instrument will prompt you to connect three standards (open,
short and load) and measure them. See
.
Figure 3-14
Open, Short, Load Connections
3. Press
after connecting each standard.
4. The analyzer will measure each standard and then calculate new
calibration coefficients. The message "
Calibration complete.
"
will appear for a few seconds when the analyzer is done calculating
the new error correction array.
CAL
Cal Kit
CAL
1-Port
Measure Standard
Summary of Contents for HP 8712ET
Page 8: ...viii ET User s Guide ...
Page 17: ...ET User s Guide 1 1 1 Installing the Analyzer ...
Page 34: ...1 18 ET User s Guide Installing the Analyzer Preventive Maintenance ...
Page 35: ...ET User s Guide 2 1 2 Getting Started ...
Page 52: ...2 18 ET User s Guide Getting Started Performing the Operator s Check ...
Page 53: ...ET User s Guide 3 1 3 Making Measurements ...
Page 118: ...3 66 ET User s Guide Making Measurements Measuring Impedance Magnitude ...
Page 119: ...ET User s Guide 4 1 4 Using Instrument Functions ...
Page 143: ...ET User s Guide 4 25 Using Instrument Functions Using Markers Figure 4 15 Delta Marker Mode ...
Page 149: ...ET User s Guide 4 31 Using Instrument Functions Using Limit Testing Figure 4 16 Limit Lines ...
Page 207: ...ET User s Guide 5 1 5 Optimizing Measurements ...
Page 226: ...5 20 ET User s Guide Optimizing Measurements Measuring Devices with Long Electrical Delay ...
Page 227: ...ET User s Guide 6 1 6 Calibrating for Increased Measurement Accuracy ...
Page 266: ...6 40 ET User s Guide Calibrating for Increased Measurement Accuracy Checking the Calibration ...
Page 267: ...ET User s Guide 7 1 7 Front Rear Panel ...
Page 292: ...7 26 ET User s Guide Front Rear Panel Line Module ...
Page 293: ...ET User s Guide 8 1 8 Hardkey Softkey Reference ...
Page 395: ...ET User s Guide 9 1 9 Specifications ...
Page 397: ...ET User s Guide 9 3 Specifications System Performance ...
Page 443: ...ET User s Guide 10 1 10 Safety and Regulatory Information ...
Page 449: ...ET User s Guide 10 7 Safety and Regulatory Information Regulatory Information ...
Page 450: ...10 8 ET User s Guide Safety and Regulatory Information Regulatory Information ...
Page 451: ...ET User s Guide 11 1 11 Factory Preset State and Memory Allocation ...