ET User’s Guide
3-15
Making Measurements
Measuring Transmission Response
Measuring Transmission Response
This section uses an example measurement to describe how to calibrate
for and make a basic transmission response measurement. This example
uses an enhanced response calibration. See the table below for a list of
all possible calibration choices for a transmission response
measurement.
TIP
An optimum calibration is critical for achieving best measurement
accuracy. Refer to
Chapter 6, “Calibrating for Increased Measurement
for detailed information about all of the various aspects of
calibration.
Table 3-2
Calibration Choices for a Transmission Response Measurement
In this example, a bandpass filter like the one that was supplied with
your network analyzer is used.
Enter the Measurement Parameters
Press
(
or
) on the analyzer
to set the analyzer to the default mode which includes measuring
transmission on measurement channel 1.
NOTE
This example measurement uses the default instrument parameters for
a transmission measurement. If your particular transmission
measurement requires specific parameters (such as frequency range,
source power level, number of data points, and sweeptime), enter them
now.
Measurement Type
Calibration Choices
Transmission
Default Response
Response
Response & Isolation
Enhanced Response
Normalize
PRESET
User Preset
Factory Preset
Summary of Contents for HP 8712ET
Page 8: ...viii ET User s Guide ...
Page 17: ...ET User s Guide 1 1 1 Installing the Analyzer ...
Page 34: ...1 18 ET User s Guide Installing the Analyzer Preventive Maintenance ...
Page 35: ...ET User s Guide 2 1 2 Getting Started ...
Page 52: ...2 18 ET User s Guide Getting Started Performing the Operator s Check ...
Page 53: ...ET User s Guide 3 1 3 Making Measurements ...
Page 118: ...3 66 ET User s Guide Making Measurements Measuring Impedance Magnitude ...
Page 119: ...ET User s Guide 4 1 4 Using Instrument Functions ...
Page 143: ...ET User s Guide 4 25 Using Instrument Functions Using Markers Figure 4 15 Delta Marker Mode ...
Page 149: ...ET User s Guide 4 31 Using Instrument Functions Using Limit Testing Figure 4 16 Limit Lines ...
Page 207: ...ET User s Guide 5 1 5 Optimizing Measurements ...
Page 226: ...5 20 ET User s Guide Optimizing Measurements Measuring Devices with Long Electrical Delay ...
Page 227: ...ET User s Guide 6 1 6 Calibrating for Increased Measurement Accuracy ...
Page 266: ...6 40 ET User s Guide Calibrating for Increased Measurement Accuracy Checking the Calibration ...
Page 267: ...ET User s Guide 7 1 7 Front Rear Panel ...
Page 292: ...7 26 ET User s Guide Front Rear Panel Line Module ...
Page 293: ...ET User s Guide 8 1 8 Hardkey Softkey Reference ...
Page 395: ...ET User s Guide 9 1 9 Specifications ...
Page 397: ...ET User s Guide 9 3 Specifications System Performance ...
Page 443: ...ET User s Guide 10 1 10 Safety and Regulatory Information ...
Page 449: ...ET User s Guide 10 7 Safety and Regulatory Information Regulatory Information ...
Page 450: ...10 8 ET User s Guide Safety and Regulatory Information Regulatory Information ...
Page 451: ...ET User s Guide 11 1 11 Factory Preset State and Memory Allocation ...