5-8
ET User’s Guide
Optimizing Measurements
Increasing Sweep Speed
To Turn Off Spur Avoidance
When spur avoidance is on (off is the factory preset default value), the
analyzer breaks each sweep into segments. Between sweep segments,
the analyzer stops and changes internal frequencies to ensure that
unwanted mixing products don’t affect the measurement. Since the
analyzer sweep is not interrupted when this feature is off, turn off spur
avoidance to increase sweep speed.
1. Press
and look at the
softkey.
2. If the
key is highlighted with a box around it, press the
key.
NOTE
If Spur Avoid must be used in your measurement, set the start frequency
as high as possible to obtain the fastest possible sweeps.
To Avoid Frequency Bandcrossings
(HP 8714ET only)
Sweep time is increased when the analyzer encounters a bandcrossing
point. The frequency bandcrossing points are approximately:
• 1900 MHz
• 2310 MHz
• 2620 MHz
Press
and then change the start frequency, stop frequency, or
span to avoid sweeping through these bandcrossing points when
possible.
MENU
Spur Avoid Options
Spur Avoid
Spur Avoid
None
FREQ
Summary of Contents for HP 8712ET
Page 8: ...viii ET User s Guide ...
Page 17: ...ET User s Guide 1 1 1 Installing the Analyzer ...
Page 34: ...1 18 ET User s Guide Installing the Analyzer Preventive Maintenance ...
Page 35: ...ET User s Guide 2 1 2 Getting Started ...
Page 52: ...2 18 ET User s Guide Getting Started Performing the Operator s Check ...
Page 53: ...ET User s Guide 3 1 3 Making Measurements ...
Page 118: ...3 66 ET User s Guide Making Measurements Measuring Impedance Magnitude ...
Page 119: ...ET User s Guide 4 1 4 Using Instrument Functions ...
Page 143: ...ET User s Guide 4 25 Using Instrument Functions Using Markers Figure 4 15 Delta Marker Mode ...
Page 149: ...ET User s Guide 4 31 Using Instrument Functions Using Limit Testing Figure 4 16 Limit Lines ...
Page 207: ...ET User s Guide 5 1 5 Optimizing Measurements ...
Page 226: ...5 20 ET User s Guide Optimizing Measurements Measuring Devices with Long Electrical Delay ...
Page 227: ...ET User s Guide 6 1 6 Calibrating for Increased Measurement Accuracy ...
Page 266: ...6 40 ET User s Guide Calibrating for Increased Measurement Accuracy Checking the Calibration ...
Page 267: ...ET User s Guide 7 1 7 Front Rear Panel ...
Page 292: ...7 26 ET User s Guide Front Rear Panel Line Module ...
Page 293: ...ET User s Guide 8 1 8 Hardkey Softkey Reference ...
Page 395: ...ET User s Guide 9 1 9 Specifications ...
Page 397: ...ET User s Guide 9 3 Specifications System Performance ...
Page 443: ...ET User s Guide 10 1 10 Safety and Regulatory Information ...
Page 449: ...ET User s Guide 10 7 Safety and Regulatory Information Regulatory Information ...
Page 450: ...10 8 ET User s Guide Safety and Regulatory Information Regulatory Information ...
Page 451: ...ET User s Guide 11 1 11 Factory Preset State and Memory Allocation ...