162
Chapter 5, HP 8924C Performance Verification
RF Gen Duplex Output High Level Accuracy Performance Test 5
N:\mkt\MANUALS\HP8924C\ALR\24c_pt.fb
NOTE:
Make sure to enter the power sensor’s calibration data into the measuring receiver and zero
the sensor module. Refer to the measuring receiver’s and the sensor module’s operating
manuals.
Procedure
1. Access the CONFIGURE screen.
2. Change the
RF Display
field to
Freq
.
3. Access the RF GENERATOR screen.
4. Set the
Output Port
field to
Dupl.
5. Set the
AFGEN1 TO
and
AFGEN2 To
fields to
Off (using the ON/OFF
key).
6. Measure the RF power at the frequencies and levels shown in the PTR (
Duplex Output High Level Accuracy Performance Test 5" on page 314
) and
compare the measured RF power to the limits shown in the PTR.
AFGen2 Freq
AFGen2 To
RF Gen Freq
Amplitude
Atten Hold
Output Port
AFGen 1 Freq
AF Gen1 To
FM Coupling
Audio Out
Mic Pre-Emp
Mod In To
To Screen
TX Frequency
TX Power
RF GENERATOR
AC Level
Level (dBm) RF Freq (MHz)
Select Off
Select Dupl
Summary of Contents for 8924E
Page 52: ...Contents 52 N mkt MANUALS HP8924C ALR 8924CE TOC A Error Messages ...
Page 53: ...Contents 53 N mkt MANUALS HP8924C ALR 8924CE TOC Index 633 ...
Page 54: ...Contents 54 N mkt MANUALS HP8924C ALR 8924CE TOC ...
Page 137: ...137 N mkt MANUALS HP8924C ALR per_adj fb 4 Periodic Adjustments ...
Page 449: ...449 N mkt MANUALS HP8924C ALR service fb 10 Service Screen ...
Page 455: ...455 N mkt MANUALS HP8924C ALR blkdia fb 11 Block Diagrams ...
Page 522: ...522 Chapter 11 Block Diagrams Assembly Details N mkt MANUALS HP8924C ALR blkdia fb ...
Page 613: ...613 N mkt MANUALS HP8924C ALR errmssg fb A Error Messages ...