Chapter 6, HP 8924E Performance Verification
RF Gen Duplex Output High Level Accuracy Performance Test 5
N:\mkt\MANUALS\HP8924C\ALR\24e_pt.fb
NOTE:
Make sure to enter the power sensor’s calibration data into the measuring receiver and zero
the sensor module. Refer to the measuring receiver’s and the sensor module’s operating
manuals.
Procedure
1. Press the PRESET key on the Test Set.
2. Access the CONFIGURE screen.
3. Change the
RF Display
field to
Freq
.
4. Set the
Output Port
field to
Dupl.
5. Access the ANALOG MEAS screen.
6. Set the
AFGEN1 TO
field to
Off (using the ON/OFF key).
7. Measure the RF power at the frequencies and levels shown in the PTR (
Duplex Output High Level Accuracy Performance Test 5" on page 366
) and
compare the measured RF power to the limits shown in the PTR.