136
Chapter 3, Repair
Replacing the A16 Host Controller
N:\mkt\MANUALS\HP8924C\ALR\repair.fb
Switch S1 Settings
The following table lists the factory default settings for the four-position DIP
switch, S1. Note that the Write Protect function is set to allow calibration data to
be entered when the automated periodic adjustments are made. To safeguard
calibration data between periodic adjustments, set Switch 1 to the OPEN position.
•
Switch 1: Write Calibration EEPROM (Open=Disabled, Closed=Enabled)
•
Switch 2: Smart Card (Open=PCMCIA, Closed=Epson)
•
Switch 3: Run from Flash/OTP (Open=Flash, Closed=OTP) Switch 4: Not
Used
Table 6
Default Controller Switch Settings
Switch 1
Switch 2
Switch 3
Switch 4
Closed
Open
Open
Not Used