301
Chapter 6, HP 8924E Performance Verification
CDMA Analyzer Calibrated Tuned Channel Power Level Accuracy Performance Test 31
N:\mkt\MANUALS\HP8924C\ALR\24e_pt.fb
Measuring Receiver
Set the display mode to LOG.
Set the measurement to Tuned RF Level.
Calibrate the power sensor.
NOTE:
The power sensor’s calibration data must be entered into the measuring receiver and the sensor
module must be zeroed. Refer to the measuring receiver’s and the sensor module’s operating
manual.
Measurement Procedure
1. Press the PRESET key on the Test Set.
2. Access the CONFIGURE screen.
3. Set the
RF Display
field to
Freq
.
4. Set the
(GEN)-(ANL)
field to
0 MHz
.
5. Change
Output Port
to
Duplex Out
.
6. Set the
Input Atten
field to
Hold
and
0 dB
.
7. Access CDMA Call Control Screen.
8. Set the
RF Gen Freq
field to
840 MHz
.
9. Measure each of the RF levels in the PTR with the Test Set and the measuring
receiver (
see "CDMA Analyzer Calibrated Tuned Channel Power Level Accuracy
Performance Test 31" on page 403
) and record the results.
Summary of Contents for 8924E
Page 52: ...Contents 52 N mkt MANUALS HP8924C ALR 8924CE TOC A Error Messages ...
Page 53: ...Contents 53 N mkt MANUALS HP8924C ALR 8924CE TOC Index 633 ...
Page 54: ...Contents 54 N mkt MANUALS HP8924C ALR 8924CE TOC ...
Page 137: ...137 N mkt MANUALS HP8924C ALR per_adj fb 4 Periodic Adjustments ...
Page 449: ...449 N mkt MANUALS HP8924C ALR service fb 10 Service Screen ...
Page 455: ...455 N mkt MANUALS HP8924C ALR blkdia fb 11 Block Diagrams ...
Page 522: ...522 Chapter 11 Block Diagrams Assembly Details N mkt MANUALS HP8924C ALR blkdia fb ...
Page 613: ...613 N mkt MANUALS HP8924C ALR errmssg fb A Error Messages ...