272
Chapter 6, HP 8924E Performance Verification
RF Analyzer Level Accuracy Performance Test 21
N:\mkt\MANUALS\HP8924C\ALR\24e_pt.fb
RF Analyzer Level Accuracy
Performance Test 21
Description
A function generator’s level is set to approximately 500 mW and measured with a
power meter. Then, the power is measured with the unit-under-test, and the two
measurements are compared.
A dual directional coupler is used to minimize mismatch uncertainty.
Calibration Setup
Function Generator
OUTPUT
INPUT
Measuring Receiver
SENSOR
Incident
Test Port
50
Ω
Termination
OUTPUT
Sensor Module
No Connection
Direct Connection (no cables)
Dual Directional Coupler