259
Chapter 6, HP 8924E Performance Verification
AF Analyzer AC Voltage Accuracy Performance Test 14
N:\mkt\MANUALS\HP8924C\ALR\24e_pt.fb
Procedure
1. Press the PRESET key on the Test Set.
2. Access the ANALOG MEAS screen.
3. Set the
AF ANL In
field to
Audio In.
4. Set
Filter 1
to
<20 Hz HPF
, and
set Filter 2
to
>99 kHz LP
.
5. Set the
De-Emphasis
field to
Off
.
6. Set the
Detector
field to
RMS
.
7. Set the audio analyzer’s source to the frequencies and levels shown in the PTR (
"AF Analyzer AC Voltage Accuracy Performance Test 14" on page 384
), adjust the
level until the DVM reads the correct level.
8. Measure the ac level on the Test Set.
9. Compare the measured ac level to the limits shown in the PTR.
Summary of Contents for 8924E
Page 52: ...Contents 52 N mkt MANUALS HP8924C ALR 8924CE TOC A Error Messages ...
Page 53: ...Contents 53 N mkt MANUALS HP8924C ALR 8924CE TOC Index 633 ...
Page 54: ...Contents 54 N mkt MANUALS HP8924C ALR 8924CE TOC ...
Page 137: ...137 N mkt MANUALS HP8924C ALR per_adj fb 4 Periodic Adjustments ...
Page 449: ...449 N mkt MANUALS HP8924C ALR service fb 10 Service Screen ...
Page 455: ...455 N mkt MANUALS HP8924C ALR blkdia fb 11 Block Diagrams ...
Page 522: ...522 Chapter 11 Block Diagrams Assembly Details N mkt MANUALS HP8924C ALR blkdia fb ...
Page 613: ...613 N mkt MANUALS HP8924C ALR errmssg fb A Error Messages ...