583
Chapter 14, Diagnostics Descriptions
Description of RF Diagnostics (RFDIAGS3)
N:\mkt\MANUALS\HP8924C\ALR\diagnos.fb
Signal Generator Synthesizer A24
Table 123
A24 Signal Generator Synthesizer Tests
Measurement
Explanation and Suggestions
Measurement Point
Reference
Detector
(Ref present)
The reference present detector is read with the 200
kHz reference signal enabled. If this measurement
fails, trace the 200 kHz signal from P3-1 of the A27
Reference to P3-3 of the A24 Signal Generator Syn-
thesizer.
Latch:gsyn_ref_not_present_sense
Out-of-Lock
Detector
(Locked)
The Out-of-Lock Detector is read with the loop set
to be locked.
Latch:gsyn_out_of_lock_int_sense
Reference
Detector
(Reference
NOT present)
The reference-present detector is read with the 200
kHz the Reference signal disabled at the Reference
assembly.
Latch:gsyn_ref_not_present_sense
Out-of-Lock
Detector
(NOT locked)
The out-of-lock detector at the Reference assembly.
This should cause the loop to be out of lock.
Latch:gsyn_out_of_lock_int_sense
Frequency
Range
The 200 kHz Reference is enabled at the Reference
assembly. Then the frequency of this loop is incrim-
inated from 500 to 1000 MHz in 50 MHz steps. The
out-of-lock detector is read at each step.
Latch:gsyn_out_of_lock_int_sense
Summary of Contents for 8924E
Page 52: ...Contents 52 N mkt MANUALS HP8924C ALR 8924CE TOC A Error Messages ...
Page 53: ...Contents 53 N mkt MANUALS HP8924C ALR 8924CE TOC Index 633 ...
Page 54: ...Contents 54 N mkt MANUALS HP8924C ALR 8924CE TOC ...
Page 137: ...137 N mkt MANUALS HP8924C ALR per_adj fb 4 Periodic Adjustments ...
Page 449: ...449 N mkt MANUALS HP8924C ALR service fb 10 Service Screen ...
Page 455: ...455 N mkt MANUALS HP8924C ALR blkdia fb 11 Block Diagrams ...
Page 522: ...522 Chapter 11 Block Diagrams Assembly Details N mkt MANUALS HP8924C ALR blkdia fb ...
Page 613: ...613 N mkt MANUALS HP8924C ALR errmssg fb A Error Messages ...