103
Chapter 3, Repair
Downloading Calibration Data
N:\mkt\MANUALS\HP8924C\ALR\repair.fb
Downloading Calibration Data
Most assemblies in the Test Set require calibration data. To ensure that the Test
Set remains calibrated after an assembly is replaced, new calibration data must be
downloaded. When required, calibration data is provided on a PCMCIA memory
card that is included with the replacement assembly. Refer to
Assemblies To Troubleshooting Aids" on page 85
, to see which modules require
calibration data.
Calibration Data Download Procedure
1. Switch the Test Set’s power off.
2. Remove the faulty assembly.
3. Install the replacement assembly.
4. Switch the Test Set’s power on.
5. Insert the memory card.
6. Press the TESTS key.
7. Set the
Select Procedure Location
: field to
Card
.
8. Set the
Select Procedure Filename
: field to:
DNLDCAL
.
9. Press the K1 key to run the test.
10. Follow the instructions on the screen.
Summary of Contents for 8924E
Page 52: ...Contents 52 N mkt MANUALS HP8924C ALR 8924CE TOC A Error Messages ...
Page 53: ...Contents 53 N mkt MANUALS HP8924C ALR 8924CE TOC Index 633 ...
Page 54: ...Contents 54 N mkt MANUALS HP8924C ALR 8924CE TOC ...
Page 137: ...137 N mkt MANUALS HP8924C ALR per_adj fb 4 Periodic Adjustments ...
Page 449: ...449 N mkt MANUALS HP8924C ALR service fb 10 Service Screen ...
Page 455: ...455 N mkt MANUALS HP8924C ALR blkdia fb 11 Block Diagrams ...
Page 522: ...522 Chapter 11 Block Diagrams Assembly Details N mkt MANUALS HP8924C ALR blkdia fb ...
Page 613: ...613 N mkt MANUALS HP8924C ALR errmssg fb A Error Messages ...