210
Chapter 5, HP 8924C Performance Verification
Spectrum Analyzer Image Rejection Performance Test 26
N:\mkt\MANUALS\HP8924C\ALR\24c_pt.fb
Image Rejection Procedure
1. Press the PRESET key on the Test Set.
2. Access the CONFIGURE screen, and set the
RF Display
field to
Freq
.
3. Access the SPECTRUM ANALYZER screen.
4. Set the
RF In/Ant
field to
Ant
.
5. Set the
Ref Level
field to
−
25 dBm
.
6. Set the
Span
field to
5 kHz
.
7. Set the signal generator’s frequency, and the spectrum analyzer’s
Center Freq
field,
to the frequencies shown in the PTR (
see "Image Rejection" on page 348
). The image
response is the spectrum analyzer’s marker level plus the signal generator’s output level
(-20 dBm).
Lvl
Freq
To Screen
Span
Ref Level
Center Freq
Controls
SPECTRUM ANALYZER
Marker
Select Ant
Enter 5 kHz
Enter -25 dBm
Set to the frequency listed in the PTR
Measured Level
Summary of Contents for 8924E