![HP 8924E Assembly Download Page 434](http://html.mh-extra.com/html/hp/8924e/8924e_assembly_166194434.webp)
434
Chapter 9, HP 8924C/E Specifications
CDMA Analyzer Specifications
N:\mkt\MANUALS\HP8924C\ALR\specs.fb
CDMA Analyzer Specifications
CDMA Average Power Measurement
To achieve the specified accuracy when measuring power at the RF IN/OUT port,
the internal signal generator’s level must be 40 dB below the measured power or
less than
−
20 dBm at the DUPLEX output port.
Input Frequency Range
•
30 MHz to 1000 MHz.
Input Connector
•
RF IN/OUT Connector Only.
Measurement Bandwidth
•
Provides an accurate measure of the total power for all present signals within ±2 MHz
of the specified operating frequency. If other signals are present outside of this
frequency range, reduced measurement accuracy will result.
Maximum Input Level
•
+35 dBm (3 Watts continuous, 10 Watts allowable for 10 seconds per minute).
Measurement Range
•
−
10 dBm to +35 dBm.