![HP 8924E Assembly Download Page 265](http://html.mh-extra.com/html/hp/8924e/8924e_assembly_166194265.webp)
265
Chapter 6, HP 8924E Performance Verification
AF Analyzer DC Level Accuracy Performance Test 17
N:\mkt\MANUALS\HP8924C\ALR\24e_pt.fb
Procedure
1. Press the PRESET key on the Test Set.
2. Access the ANALOG MEAS screen and set the
AF Anl In
field to
Audio In
.
3. Set the lower-right measurement to
DC Level
.
4. Set the
AFGen1 Freq
field to
0 kHz.
5. Set the
AFGen1 To
field to
Audio Out
.
6. Set the
AFGen1
level to the levels shown in the PTR (
Accuracy Performance Test 17" on page 388
), adjust until the voltmeter reads the
correct level.
7. Read the DC Level and compare the reading to the limits shown in the PTR.