![HP 8924E Assembly Download Page 269](http://html.mh-extra.com/html/hp/8924e/8924e_assembly_166194269.webp)
269
Chapter 6, HP 8924E Performance Verification
AF Analyzer Frequency Accuracy at 400 kHz Performance Test 19
N:\mkt\MANUALS\HP8924C\ALR\24e_pt.fb
Procedure
1. Press the PRESET key on the Test Set.
2. Access the ANALOG MES screen, and set the
AF Anl In
field to
Audio In
.
3. Set the
Filter 2
field to
>99 kHz LP
, and the lower-right measurement field to
AF Freq.
4. Read the
AF Freq
on the Test Set and compare the reading to the limits shown in the
PTR (
see "AF Analyzer Frequency Accuracy at 400 kHz Performance Test 19" on
Summary of Contents for 8924E