Chapter 7, HP 8924C Performance Test Records
AF Analyzer Distortion, SINAD and SNR Accuracy Performance Test 16
N:\mkt\MANUALS\HP8924C\ALR\24c_ptr.fb
AF Analyzer Distortion, SINAD and SNR Accuracy Performance Test 16
Table 28
AF Analyzer Distortion and SINAD Accuracy
Harmonic
Frequency
(kHz)
Harmonic
Level (V)
Measurement
Type
Distortion (%) and SINAD (dB) Accuracy
Limits
Lower
Upper
Actual
2
.1
Distortion
8.856
11.144
2
.1
SINAD
19.043
21.043
3
.1
Distortion
8.856
11.144
3
.1
SINAD
19.043
21.043
2
.01
Distortion
.890
1.120
2
.01
SINAD
39.000
41.000
3
.01
Distortion
.890
1.120
3
.01
SINAD
39.000
41.000
2
.005
Distortion
.445
.560
2
.005
SINAD
45.021
47.021
3
.005
Distortion
.445
.560
3
.005
SINAD
45.021
47.021