221
Chapter 5, HP 8924C Performance Verification
CDMA Analyzer Average Power Level Accuracy Performance Test 30
N:\mkt\MANUALS\HP8924C\ALR\24c_pt.fb
Make the following signal generator setting:
•
Amplitude 0 dB
Make the following measuring receiver settings:
•
Track mode
•
RF Power measurement
•
4.4 SPCL
•
8.3 SPCL
Calibration Procedure
1. Set the signal generator and the measuring receiver to 40 MHz.
2. Make an RF POWER measurement, in mW, with the measuring receiver. Record the
measured power as Reference Power in the Measurement Procedure in step 11.
3. Disconnect the sensor cable at the measuring receiver. (leave the sensor connected to
the directional coupler so that it will be terminated into 50
Ω
4. Make a TUNED RF LEVEL measurement, in mW, with the measuring receiver.
Record the measured level as Tuned RF #1 in the Measurement Procedure in step 11.
5. Repeat this procedure for each of the frequencies listed in the PTR (
Analyzer Average Power Level Accuracy Performance Test 30" on page 353
).