Chapter 4 Command Reference
Measurement Instructions (:CONFigure, :FETCh, :MEASure, :READ)
4-74
Programming Guide
Descriptions of the Measurement Functions— <function> (Cont.)
:MEASure[:SCALar][:VOLTage]:TOTalize:TIMed? [<gate_time>][,(@1)]
Measures Totalize during the specified <gate time>.
The totalize gate mode is set to “time,” and the totalize gate time is defaulted or set
to the specified value.
The trigger settings are coupled so that the measurement channel has auto-trigger
disabled with a positive slope.
<gate_time>
range:
For short gate time: 100E-5 to 999E-5 seconds
For long gate time: 10E-3 to 1000.000 seconds
resolution:
For short gate time: 1E-5 seconds
For long gate time: 1E-3 seconds
default:
.100 second
<source_list>
range:
(@1)
default:
(@1)