Chapter 2 Command Summary
HP 53131A/132A Command Summary
2-28
Programming Guide
Table 2-2. HP 53131A/132A SCPI Command Summary (Continued)
Keyword/Syntax
Parameter Form
Std/
New
Comments
:DIAGnostic (Cont.)
:CALibration (Cont.)
:ROSCillator
:AUTO
:STATus?
:TINTerval
:FINE
:QUICk
ONCE | OFF
[1 | 2 | 3 | 4]
New
New
New
New
New
New
Subtree. ROSCillator is an abbreviation for
Reference OSCillator.
ONCE calibrates the timebase. This command is
usable only if the instrument contains the medium or
high stability oscillator option.
Query only. Returns status of last calibration.
0 = pass; 1 = fail.
Subtree.
Event; no query. Four steps which calibrate out
Ch1, 2 electrical path length differences.
Event; no query. Calibrates out Ch1, 2 electrical path
length differences.
:DISPlay
:ENABle
:MENU
[:STATe]
[:WINDow]
:TEXT
:FEED
:RADix
<Boolean>
OFF
“[:]CALCulate2 ” |
“[:]CALCulate3 ”
COMMa | DPOint
Std
Std
Std
Std
Std
Std
Std
New
Subsystem. Controls the selection and presentation
of textual information on the display.
Controls whether the whole display is visible.
Subtree.
Sets the Counter to switch from the menu display to
the result display.
Subtree.
Subtree. Allows for the display of textual information.
Sets which data flow is fed into the display. “CALC2”
specifies the raw measurement, scaled/offset
measurement, or Limit Graph display. “CALC3”
specifies the statistics result display.
Sets the character used to separate integral and
fractional portions of a number. (USA numerical
convention is Decimal POint.)
:FETCh
Std
See Measurement Instructions in this table.
:FORMat
[:DATA]
ASCii | REAL
Std
Std
Subsystem. Sets a data format for transferring
numeric information.
Sets the data format.
:HCOPy
:CONTinuous
<Boolean>
New
New
Enables or disables printing results.
Summary of Contents for 53131A
Page 1: ...Programming Guide HP 53131A 132A 225 MHz Universal Counter ...
Page 2: ......
Page 13: ...1 1 Before You Start ...
Page 24: ...Chapter 1 Before You Start Related Documentation 1 12 Programming Guide ...
Page 25: ...2 2 Command Summary A Quick Reference ...
Page 68: ...Chapter 2 Command Summary RST Response 2 44 Programming Guide ...
Page 69: ...3 3 Programming Your Universal Counter for Remote Operation ...
Page 167: ...4 4 Command Reference A Dictionary ...
Page 310: ...Chapter 4 Command Reference WAI Wait to Continue Command 4 144 Programming Guide ...
Page 311: ...5 5 Errors ...
Page 329: ......