Chapter 4 Command Reference
Introduction
Programming Guide
4-29
•
The calibration state is stored in non-volatile memory, and is unaffected by
power-on, save/recall, and *RST.
•
The security code is set to 53131 or 53132 (depending on which model you
have) when the Counter is shipped from the factory.
If you forget your security code, you can reset the security code to the model-
number default by resetting all of the non-volatile memory to a default state.
See the Assembly-Level Service Guide for more information.
Scale & Offset / POWER (Calibration Menu)
Comments
Related
Front-Panel
Keys