Chapter 4 Command Reference
Introduction
Programming Guide
4-121
:TRIGger Subsystem
This subsystem enables synchronization of instrument actions with specified
internal or external events.
:TRIGger:COUNt:AUTO <Boolean>
Sets or queries the control over the number of measurements made when
:INITiate[:IMMediate] is performed.
When :TRIG:COUN:AUTO is OFF, then :INIT[:IMM] initiates a single
measurement.
When :TRIG:COUN:AUTO is ON and [:SENS]:FUNC[:ON] is not Totalize or
Voltage Peak function, then:
•
when statistics are enabled (:CALC3:AVER[:STAT] is ON), :INIT[:IMM]
will initiate a complete block of N (:CALC3:AVER:COUN) valid
measurements.
•
when statistics are enabled (:CALC3:AVER[:STAT] is ON),
and limit-filtering is enabled (:CALC3:LFIL:STAT is ON), :INIT[:IMM] will
initiate a complete block of N (:CALC3:AVER:COUN) valid, in-limit
measurements.
•
when statistics are disabled (:CALC3:AVER[:STAT] is OFF), :INIT[:IMM]
initiates a single measurement.
•
Single ASCII-encoded byte, 0 or 1.
•
A value of 0 indicates OFF; a value of 1 indicates ON.
•
*RST: OFF
•
The :TRIG:COUN:AUTO setting has no affect on the operation of
:INIT:CONT ON, which always behaves as if :TRIG:COUN is set to 1.
Stats
Query Response
Comments
Related
Front-Panel
Keys
Summary of Contents for 53131A
Page 1: ...Programming Guide HP 53131A 132A 225 MHz Universal Counter ...
Page 2: ......
Page 13: ...1 1 Before You Start ...
Page 24: ...Chapter 1 Before You Start Related Documentation 1 12 Programming Guide ...
Page 25: ...2 2 Command Summary A Quick Reference ...
Page 68: ...Chapter 2 Command Summary RST Response 2 44 Programming Guide ...
Page 69: ...3 3 Programming Your Universal Counter for Remote Operation ...
Page 167: ...4 4 Command Reference A Dictionary ...
Page 310: ...Chapter 4 Command Reference WAI Wait to Continue Command 4 144 Programming Guide ...
Page 311: ...5 5 Errors ...
Page 329: ......
Page 330: ...H Manual Part Number 53131 90044 Printed in U S A NOVEMBER 1996 ...