Chapter 4 Command Reference
Measurement Instructions (:CONFigure, :FETCh, :MEASure, :READ)
4-78
Programming Guide
The work-around commands which should be substituted are listed below.
:INIT; *WAI
:FETCh:MAXimum?
:INIT; *WAI
:FETCh:MINimum?
:INIT; *WAI
:FETCh:PTPeak?
The command listed below should NOT be used when the function choice is
Voltage Peaks.
:READ?
The work-around command sequence which should be substituted is listed below.
:INIT; *WAI
:FETCh?
When the function is set to Voltage Peaks, and the Counter is in SINGLE mode,
the HP-IB command :FETCh? will hang the HP-IB if the measurement has not
completed.
The work around inserts a command which enforces correct timing. The
work-around command sequence is listed below.
:INIT; *WAI
:FETCh?
When the function is set to Voltage Peaks, and the Counter is in RUN mode, the
HP-IB command :FETCh? should not be used.
The work around puts the box into SINGLE mode, and inserts a command which
enforces correct timing. The work -around command sequence is listed below.
:INIT:CONTinuous OFF
:INIT; *WAI
:FETCh?
Summary of Contents for 53131A
Page 1: ...Programming Guide HP 53131A 132A 225 MHz Universal Counter ...
Page 2: ......
Page 13: ...1 1 Before You Start ...
Page 24: ...Chapter 1 Before You Start Related Documentation 1 12 Programming Guide ...
Page 25: ...2 2 Command Summary A Quick Reference ...
Page 68: ...Chapter 2 Command Summary RST Response 2 44 Programming Guide ...
Page 69: ...3 3 Programming Your Universal Counter for Remote Operation ...
Page 167: ...4 4 Command Reference A Dictionary ...
Page 310: ...Chapter 4 Command Reference WAI Wait to Continue Command 4 144 Programming Guide ...
Page 311: ...5 5 Errors ...
Page 329: ......
Page 330: ...H Manual Part Number 53131 90044 Printed in U S A NOVEMBER 1996 ...